Applied Science and Convergence Technology 2020; 29(4): 77-81
Published online July 31, 2020
https://doi.org/10.5757/ASCT.2020.29.4.077
Copyright © The Korean Vacuum Society.
Hee Jung Yeoma , b , Jang Jae Leec , Si Jun Kima , Young Seok Leea , Chul hee Choa , Jung Hyung Kimb , Hyo-Chang Leeb , and ShinJae Youa , *
aDepartment of Physics, Chungnam National University, Daejeon 34134, Republic of Korea
bKorea Research Institute of Standards and Science, Daejeon 34113, Republic of Korea
cInstitute of Quantum System (IQS), Chungnam National University, Daejeon 34134, Republic of Korea
Correspondence to:E-mail: sjyou@cnu.ac.kr
This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-CommercialLicense (http://creativecommons.org/licenses/by-nc/3.0) which permits unrestricted non-commercial use, distribution,and reproduction in any medium, provided the original work is properly cited.
In this study, we confirmed through experiments that a noise signal appears in the low-mass region when measuring plasma using a quadrupole mass spectrometer. We conducted a numerical analysis to identify the cause of this occurrence. It was determined that high-energy ions generated in plasma are not filtered in the low quadrupole filter voltage region owing to the high velocity of the ions. We thus detected the signal in the low-mass region based on the stability area. In this paper, we identify through numerical analysis the quadrupole filter length and the radius of the incident ion beam that effectively reduce low-mass noise.
Keywords: Quadrupole mass filter, Radical, Plasma
Recently, it has become very important to measure the gas composition and the partial pressure in industrial semiconductor and display processing, such as etching, deposition, and ashing [1–5]. Among many possible diagnostic methods, the mass spectrometer is an investigative tool that offers many advantages for analyzing reactive plasmas on account of its high sensitivity and mass resolution characteristics [3,6–9]. Many kinds of mass spectrometers exist, such as the time of flight mass spectrometer [10], magnetic sector mass spectrometer [11,12], and quadrupole mass spectrometer (QMS) [13]. Among these, QMS is often used for measuring radicals in plasma owing to its low cost, relatively small size, and robustness [8,14]. When measuring radical species in plasma using QMS, potential problems may arise, such as low mass tailing [15], changed acceptance [16,17], and abnormal signals in the low-mass region [18]. Nevertheless, previous research has not determined whether such a problem actually occurs in plasma measurement.
In this study, we confirmed that an abnormal signal occurs in the low-mass region during plasma measurement. The cause of this occurrence was determined by numerical analysis. Through a numerical simulation of the quadrupole mass spectrometer, the transmission characteristics of the ions entering the quadrupole filter in accordance with the initial conditions, such as the preliminary velocity and initial position, were analyzed. Accordingly, the design conditions for the quadrupole filter length and ion beam radius in plasma measurement were studied.
A radical measurement was performed in capacitively coupled plasma. The chamber geometry, radio-frequency power feeding, and QMS (Stanford Research Systems, RGA 200) are shown in Fig. 1. The chamber is a cylindrical shape with an inner diameter of 340 mm and a distance of 50 mm from the top to the bottom of the electrode. A 13.56 MHz power supply is connected to the bottom electrode through an L-type auto-matcher. The input power during the experiment was 0 and 500 W. The gases utilized in the experiment were Ar and C4F8 with a fixed 20 sccm flow rate, respectively. A rotary pump and turbo-molecular pump were employed to maintain the background gas pressure of the processing chamber below 5×10-6 Torr, which was measured using a Penning gauge. The experimental pressure was 10 mTorr. It was measured using a capacitive diaphragm gauge. The base pressure of the QMS was set below 2×10-7 orr using a 90 μm diameter orifice and a differential pumping system.
Figure 2 shows the mass spectra of the Ar and C4F8 gas at the input power of 0 and 500 W, and the gas pressure of 10 mTorr. Figures 2(a) and 2(b) indicate that abnormal signals (0 to 10 amu) in the low-mass region are measured under the conditions of 500 W, an Ar mass of 40 amu, and a carbon mass of 12 amu. The carbon mass is the smallest element that can be measured by a dissociation in C4F8 plasma. Therefore, the signal measured at the low mass is not that of the elements generated by plasma. According to Medley
The QMS contained three basic elements: an ion source, a quad upole mass filter (QMF), and an ion detector. The electrode structure of the QMF is shown in Fig. 3, where ϕ0 is the applied electrode potential,
The equations of motion are
Figure 4 shows Mathieu’s first stability region for the (
In this simulation, we assume that the ions are monovalent, the initial ion condition has no x and y velocities, and the z velocity is incident at a velocity accelerated by the sheath potential. The initial ion velocity is calculated as
In this simulation, Eq. (3) is calculated in two steps using the Euler method [24]. First, the velocity term is given by
Then, the position term is given by
Figure 5 depicts the ion transmittance of the hydrogen ions (1amu) in accordance with the ion velocity. If the ion passing time of the quadrupole mass filter (τion) is one cycle of the RF period (1TRF), all hydrogen ions pass through in 0 to 2 amu, and an average of 72 % of the ions passes through in 2 to 4 amu. This phenomenon decreases as the speed of the ions decreases, as shown in Figs. 5(b) and 5(c). As the τion = 47TRF condition, the hydrogen atoms are filtered in the mass area except for 1 amu. In the case of plasma, it is thought that this phenomenon may occur because high-energy ions enter QMS on account of acceleration in the sheath potential [20]. This can be also confirmed in the stability region, as shown in Fig. 6. If the filter pass time of the ions is short, it can be observed that there is a point overlapping the scan line in the small RF and DC voltage regions. This means that ions reach the detector before being filtered in response to the quadrupole electric field.
Figure 7 shows the ions passing through each initial ion position according to the filter pass time at the filter voltage of
To effectively remove this low mass noise, it is necessary to adjust the QMF length with consideration of the τion and the radius of the incident ion beam. Figure 8 shows the signal ratio of the hydrogen atom to the low-mass noise according to τion for the hydrogen atom. It additionally depicts the required QMF length for the ion energy of various masses for 1 to 100 amu atoms. For quadrupole mass spectro- meters, the low-mass noise and mass signal can be distinguished if the signal at 0.5 amu is reduced to less than 10 %. This is because the mass resolution is usually less than one [4,9]. Therefore, we assume that the point at which the transmittance at 0.5 amu becomes 10 % or less of the transmittance at the hydrogen atom mass is the point at which low-mass noise disappears. In Fig. 8(a), it is observed that the ratio of 1 amu signal (
Therefore, to measure the hydrogen ions in the plasma, the length of the quadrupole filter must satisfy the τion > 11TRF condition to exclude the low-mass noise. In addition, τion = 11TRF is defined as τ
Figure 8(b) shows
Figure 9 shows the normalized signal ratio of the original mass to noise for particles with 1, 40, and 70 amu, as well as the radius of the ion beam. In the mass region of 1 to 70 amu, low mass noise is effectively removed when the beam radius is 0.2 times the filter system radius (
In this study, we analyzed the low-mass noise that can occur when measuring plasma using a quadrupole mass spectrometer. We confirmed that the abnormal signal can be caused by ions accelerated by the sheath potential. In addition, we confirmed that if the filter length is small, it can be caused by a small potential. To solve this issue, we verified that the intensity of the abnormal signal can be reduced by considering the filter length for ion energy and the diameter of the incident ion beam.
This research was supported by a National Research Council of Science & Technology (NST) grant by the Korea government (MSIP) (No. CAP-17-02-NFRI), by the Korea Institute of Energy Technology Evaluation and Planning (KETEP), and the MOTIE of the Republic of Korea (No. 20172010105910). In addition, it was supported by a Korea Institute for Advancement of Technology (KIAT) grant funded by the Korea Government (MOTIE) (P0008458, The Competency Development Program for Industry Specialist), by the Aerospace Low Observable Technology Laboratory Program of the Defense Acquisition Program Administration and the Agency for Defense Development of the Republic of Korea, and by the Basic Science Research Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Education (NRF-2020R1A6A1A03047771).